GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 70
Attribute Name Tag Type Notes
Overlay Number (0020,0022) 2 The image number of the image this overlay
is for
Overlay Date (0008,0024) 3 Date of overlay
Overlay Time (0008,0034) 3 Time of overlay
A.5.16
Requested Procedure Module
Attribute Name Tag Type Notes
Requested Procedure ID (0040,1001) 1C Sent if present in image.
Referenced Study Sequence (0008,1110) 3 Sent if present in image.
>Referenced Study Sequence Class
UID
(0008,1150) 3 Sent if present in image.
>Referenced Study Sequence
Instance UID
(0008,1155) 3 Sent if present in image.
A.5.17
Scheduled Procedure Step Module
Attribute Name Tag Type Notes
Scheduled Procedure Step
Description
(0040,0007) 1C Sent if present in image.
Scheduled Procedure Action Item
Code Sequence
(0040,0008) 3 MR does not use
Schedule Procedure Step ID (0040,0009) 1C Sent if present in image.