GE Medical Systems SIGNA PROFILE CONFORMANCE STATEMENT
REV 2.7 sm 2303347
doc ver 2.7 60
*Software Versions ( 1800, 1020 ) Printer shall return Value
*Date Last Calibrated ( 1800, 1200 ) Printer shall return Value
*Time Last Calibrated ( 1800, 1201 ) Printer shall return Value
The Print SCU issues the N-GET service to retrieve the printer status. The statuses are
processed as follows:
If Printer status (2110,0010) is NORMAL, the film is printed.
If Printer status (2110,0010) is FAILURE the print job is terminated. The Printer Status
Info (2110,0020) attribute is not processed.
If Printer status (2110,0010) is WARNING one of three things will happen:
a) If the Printer Status Info (2110,0020) is SUPPLY LOW the status is displayed to
the user and the print job continues.
b) If the Printer Status Info (2110,0020) is RECEIVER FULL or SUPPLY
EMPTY, or FILM JAM the status is displayed to the user and the print job is
aborted.
c) For all other Printer Status Info (2110,0020) values, the status is ignored and the
print job continues.
4.2.4 Association Acceptance Policy
The Print SCU does not accept associations.
4.3 COMMUNICATION PROFILES
4.3.1 Supported Communication Stacks (parts 8,9)
DICOM Upper Layer (Part 8) is supported using TCP/IP.
4.3.2 TCP/IP Stack
The TCP/IP stack is inherited from the UNIX Operating System.
4.3.2.1 API
Not Applicable
4.3.2.2 Physical Media Support
Ethernet v2.0, IEEE 802.3.